Line 20: Line 20:
   - Check vacuum on LE extension, temperature gauge on cryopump should be at minimum value. ​   - Check vacuum on LE extension, temperature gauge on cryopump should be at minimum value. ​
   - Open source gate valve by pressing either button on gate valve control at source gate.   - Open source gate valve by pressing either button on gate valve control at source gate.
 +  - Set cesium reservoir controller temperature set point (green numbers): ​ 123C for ones of uAs, 130C for tens of uAs.
   - Remove grounding hook from power supply platform and hang on gate before closing. ​   - Remove grounding hook from power supply platform and hang on gate before closing. ​
   - Close both gates to source area.  Orange lights on Glassman supplies should come on at gate closing. ​ (If not, check other gate on the other side of the accelerator;​ both must be secured.) ​   - Close both gates to source area.  Orange lights on Glassman supplies should come on at gate closing. ​ (If not, check other gate on the other side of the accelerator;​ both must be secured.) ​
Line 28: Line 29:
   - Turn up the **ionizer** heater. ​  The ionizer operates at 21-22 amps.  Keep in mind that the current will fall slightly as the ionizer gets warmer. ​ Keep adjusting the current until the ionizer is stable at 21-22 A. It should take about 1 minute to reach 21 A from the standby value of 10 A.   - Turn up the **ionizer** heater. ​  The ionizer operates at 21-22 amps.  Keep in mind that the current will fall slightly as the ionizer gets warmer. ​ Keep adjusting the current until the ionizer is stable at 21-22 A. It should take about 1 minute to reach 21 A from the standby value of 10 A.
   - Turn up the Cesium **feed tube** heater to 0.450 amperes.  ​   - Turn up the Cesium **feed tube** heater to 0.450 amperes.  ​
-  - Set the cesium **reservoir** heater to 0.260. This should be enough to give ones of microamps beam for p and d.    A setting of about 0.270 - 0.275 should produce tens of microamps. The reservoir current should not exceed 0.280.+  - Set the cesium **reservoir** heater ​current ​to **0.315** Amps. This is an approximate value above the needed power level 
   - Set the Glassman **extractor** power supply to 12.5 KV. Check to be sure the current is less than about 5 milliamps and the "​control milliamperes"​ light is NOT lit. High current indicates that the source needs to be serviced.   - Set the Glassman **extractor** power supply to 12.5 KV. Check to be sure the current is less than about 5 milliamps and the "​control milliamperes"​ light is NOT lit. High current indicates that the source needs to be serviced.
   - Set the Glassman **target** power supply to 6.5 kV. Check to be sure the current is less than about 5 milliamps and the "​control milliamperes"​ light is NOT lit. High current indicates that the source needs to be serviced. ​ A newly installed pellet may cause instability for the first hour or so, but will become stable.   - Set the Glassman **target** power supply to 6.5 kV. Check to be sure the current is less than about 5 milliamps and the "​control milliamperes"​ light is NOT lit. High current indicates that the source needs to be serviced. ​ A newly installed pellet may cause instability for the first hour or so, but will become stable.
Line 51: Line 52:
 ** Shut Down Procedure ** ** Shut Down Procedure **
  
-  - Turn down the cesium ​reservoir ​heater ​power supply.  ​Off is about 0.070 A+  - **Cesium ​reservoir**: The reservoir has a temperature controller that turns the power on/off as needed for a constant temperature.  ​The controller ​is located next to the analog temperature meter on the source power platform. ​ The red numbers are the measured temperature,​ the green numbers are the set point temperature,​ there is a small red light that indicates when the controller is calling for power. ​ The reservoir current switches between ​0.150A and .315A as directed by the controller. 
-  - Wait for the reservoir temperature to drop below 80° C.  ​+  - Wait until the reservoir controller'​s red power light is on and you can see about .315A reservoir current on the meter, run the reservoir current down on the digital meter. ​ Make sure that you see controller red light on with minimal amps of current
 +  - Wait for the reservoir temperature ​(red numbers on controller) ​to drop below** 85**° ​C.  ​
   - Turn down the feed tube power supply.   - Turn down the feed tube power supply.
   - Wait for the feed tube temperature to drop below 80° C.   - Wait for the feed tube temperature to drop below 80° C.